109
Picture 2.13-18
RUN Pass/Fail test, please follow next page:
1) Press UTILITY button to enter the “Utility menu”.
2) Press the “Next Page Page1/4” option button.
3) Press the “Next Page Page2/4” option button to enter the third page of
“Utility menu”
4) Press the “Pass/Fail” option button to enter the “Pass/Fail” function menu.
5) Press the “Enable Test” option button to select “On”.
6) Press the “Source” option button to select input signal channel.
7) Press the “Next Page Page1/2”option button to enter the second page of
the “Pass/Fail menu”.
8) Press the “Mask Setting” option button to enter the first page of “Mask
menu”.
9) Press the “X Mask” button; turn the “Universal” knob to adjust the
horizontal clearance.
10) Press the “Y Mask” button; turn the “Universal” knob to adjust the vertical
clearance.
11) Press the “Create Mask” button to create mask and you also can enter
the next page of “Mask menu” to recall the stored mask.
12) Enter the second page of “Pass/Fail function menu”, press the “Output”
Przeglądanie stron 116