39
MemDepth Detailed Specification
Memory
Depth
Channel Mode
Samping
Rate
Short
memory
Long Memory
Single
Channel
1Gsa/s 40kpts No Support
Single
Channel
500MSa/s
or lower
20kpts 2Mpts
Double
Channels
500MSa/s
or lower
20kpts 1Mpts
2.7.1 Horizontal control knob
You can use the horizontal controls to change the horizontal scale and position of
waveforms. The horizontal position readout shows the time represented by the center of
the screen, using the time of the trigger as zero. Changing the horizontal scale causes
the waveform to expand or contract around the screen center.
Horizontal “POSITION” knob
1. Adjust the horizontal position of all channels and math waveforms (the position of the
trigger relative to the center of the screen). The resolution of this control varies with the
time base setting.
2. When you press the horizontal “POSITION” Knob, you can set the horizontal position
to zero.
“S/div” knob
1. Using to change the horizontal time scale to magnify or compress the waveform. If
waveform acquisition is stopped (using the RUN/STOP or SINGLE button), turn the
S/div knob to expand or compress the waveform.
2. Select the horizontal time/div (scale factor) for the main or the window time base.
When Window Zone is enabled, it changes the width of the window zone by changing
the window time base.
Display scan mode
When the SEC/DIV control is set to100 ms/div or slower and the trigger mode is set to
Auto, the oscilloscope enters the scan acquisition mode. In this mode, the waveform