135
Appendix A: Specifications
All specification applies to 10X probe and All the SDS1000 Series Digital
Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope
must first meet the following conditions:
●
The oscilloscope must have been operating continuously for thirty
minutes within the specified operating temperature.
●
You must perform the Do Self Cal operation, accessible through the
Utility menu, if the operating temperature changes by more than 5° C.
●
The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Specifications
Inputs
Input Coupling AC, DC, GND
Input Impedance
1MΩ±2% || 16Pf±3Pf,
50Ω+/-2%(SDS1302CE/SDS1304CF/CFL,
SDS1202CE/SDS1204CF/CFL contain this function)
Maximum input voltage
400V (DC+AC PK-PK,1MΩ input impedance,
X10),CAT I
5Vrms;+/-10V Peak for 1S pulse, 50% duty cycle(50Ω
BNC) ( SDS1302CE/SDS1304CF/CFL ,
SDS1202CE/SDS1204CF/CFL contain this function)
Ch to Ch Isolation
(Both channels in same
V/div setting)
>100:1 at 150MHz: (SDS1302CE/SDS1304CF/CFL)
>100:1 at 100MHz: (SDS1202CE/SDS1204CF/CFL)
>100:1 at 70MHz (SDS1152CM)
>100:1 at 50MHz: (SDS1102C/CM/CE,
SDS1104CF/CFL)
>100:1 at 30MHz: (SDS1062C/CM/CE,
SDS1064CF/CFL)
>100:1 at 20MHz (SDS1042C)
>100:1 at 10MHz(SDS1022C)
Probe attenuator
1X、10X