SIGLENT SDS1000D Instrukcja Użytkownika Strona 143

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 142
135
Appendix A: Specifications
All specification applies to 10X probe and All the SDS1000 Series Digital
Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope
must first meet the following conditions:
The oscilloscope must have been operating continuously for thirty
minutes within the specified operating temperature.
You must perform the Do Self Cal operation, accessible through the
Utility menu, if the operating temperature changes by more than C.
The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Specifications
Inputs
Input Coupling AC, DC, GND
Input Impedance
1MΩ±2% || 16Pf±3Pf,
50+/-2%(SDS1302CE/SDS1304CF/CFL,
SDS1202CE/SDS1204CF/CFL contain this function)
Maximum input voltage
400V DC+AC PK-PK1MΩ input impedance
X10),CAT I
5Vrms+/-10V Peak for 1S pulse, 50% duty cycle(50
BNC) SDS1302CE/SDS1304CF/CFL
SDS1202CE/SDS1204CF/CFL contain this function
Ch to Ch Isolation
(Both channels in same
V/div setting)
>100:1 at 150MHz: (SDS1302CE/SDS1304CF/CFL)
>100:1 at 100MHz: (SDS1202CE/SDS1204CF/CFL)
>100:1 at 70MHz (SDS1152CM)
>100:1 at 50MHz: (SDS1102C/CM/CE,
SDS1104CF/CFL)
>100:1 at 30MHz: (SDS1062C/CM/CE,
SDS1064CF/CFL)
>100:1 at 20MHz (SDS1042C)
>100:1 at 10MHzSDS1022C
Probe attenuator
1X10X
Przeglądanie stron 142
1 2 ... 138 139 140 141 142 143 144 145 146 147 148 ... 153 154

Komentarze do niniejszej Instrukcji

Brak uwag