2.8.5 Slope & Level
The Slope and Level controls help to define the trigger. The Slope option (Edge trigger
type only) determines whether the oscilloscope finds the trigger point on the rising or the
falling edge of a signal.
The TRIGGER LEVEL knob controls where on the edge the trigger point occurs
.
Falling edge
Rising edge
T
r
i
g
g
e
r
l
e
v
e
l
c
a
n
b
e
a
d
j
u
s
t
e
d
v
e
r
t
i
c
a
l
l
y
Picture 2.8-10
2.8.6 Trigger Holdoff
You can use the Trigger Holdoff function to produce a stable display of complex
waveforms. Holdoff is time between when the oscilloscope detects one trigger and when
it is ready to detect another. The oscilloscope will not trigger during the holdoff time. For
a pulse train, you can adjust the holdoff time so the oscilloscope triggers only on the first
pulse in the train.
Note:
Press the SINGLE button when you want the oscilloscope to
acquire a single waveform.
. Trigger coupling affects only the signal passed to the trigger
system. It does not affect the bandwidth or coupling of the
signal displayed on the screen.
Normal Polarity Sync triggers always occur on negative-going
horizontal sync pulses. If the video waveform has positive-going
horizontal sync pulses, use the Inverted Polarity selection.